INSTALLATION (continued)
Fitting Selection: Check the part number to determine the required fitting mount size and thread type.
LVU800 Series is commonly installed in tank adapters, flanges, brackets or standpipes. Note: Always include
the gasket when installing the LVU800 Series.
1. Tank Adapter: Select a tank adapter fitting, such as a 1" adapter for the LVU809 series or a 2" adapter
for the LVU816, LVU826 & LVU832 series.
a. For best results, select a 2" tank adapter and add a 2" x 1" reducer bushing.
b. Avoid tank adapter (thread x thread) styles and/or pipe stops forward of the installed
transducer.
c. Always mount the tank adapter so the majority of fitting is outside the tank.
i. Never mount the tank adapter upside down or the bulk of the material is inside the
2" Tank Adapter
Socket x Thread
(LVU800‐2B shown)
2. Riser: Installations with tall, narrow risers can impede the acoustic signal.
a. LVU816, LVU826 & LVU832 Series: 2" (5 cm) diameter risers should be no taller than 5" (12.7
cm). Larger diameter risers should be no taller than 12" (30.5 cm).
b. LVU809 Series:
Note: Do not exceed the dimensions listed above
tank.
Tank Adapter (LVU800‐2B)
(LVU800‐2N80 shown)
w/ Reducer Bushing
17
2" Tank Adapter
Thread x Thread
Do not use thread x thread