GETTING STARTED (continued)
How to set using Target Calibration (Empty):
How to set using Target Calibration (Full):
Note: When setting either the TG EMPTY or TG FULL, make sure the reflective surface is perpendicular to the
LVU800 Series and not at an angle. Be careful if the tank has a sloped or cone bottom and the bottom of the
tank is exposed to air.
Instead of measuring for Empty tank, the Empty distance can be set
automatically. This method requires that the tank be Empty or at the
level that is considered Empty in the application. To set TG Empty,
follow the instructions below.
1. Before beginning, make sure the level is tank is at the Empty
level.
2. In MENU mode, select the TG CAL item.
3. When EMPTY appears, press SELECT. This sets the current
distance as the new Empty setting.
4. When EXIT appears, press SELECT return to Top Level Menu.
Instead of measuring for Full tank, the Full distance can be set
automatically. This method requires that the tank be Full or at the
level that is considered Full in the application. To set TG Full, follow
the instructions below.
1. Before beginning, make sure the level is tank is at the Full level.
2. In MENU mode, select the TG CAL item.
3. When FULL appears, press SELECT. This sets the current
distance as the new Full setting.
4. When EXIT appears, press SELECT return to Top Level Menu.