GETTING STARTED (continued) 
How to set using Target Calibration (Empty): 
 
 
How to set using Target Calibration (Full): 
 
 
 
Note: When setting either the TG EMPTY or TG FULL, make sure the reflective surface is perpendicular to the 
LVU800 Series and not at an angle.  Be careful if the tank has a sloped or cone bottom and the bottom of the 
tank is exposed to air. 
 
 
Instead  of  measuring  for  Empty  tank,  the  Empty  distance  can  be  set 
automatically.  This method requires that the tank be Empty or at the 
level  that  is  considered  Empty  in  the  application.    To  set  TG  Empty, 
follow the instructions below. 
1. Before  beginning,  make  sure  the  level  is  tank  is  at  the  Empty 
level. 
2. In MENU mode, select the TG CAL item. 
 
3. When  EMPTY  appears,  press  SELECT.    This  sets  the  current 
distance as the new Empty setting. 
4. When EXIT appears, press SELECT return to Top Level Menu. 
Instead  of  measuring  for  Full  tank,  the  Full  distance  can  be  set 
automatically.    This  method  requires  that  the  tank  be  Full  or  at  the 
level that is considered Full in the application.  To set TG Full, follow 
the instructions below. 
1. Before beginning, make sure the level is tank is at the Full level.
2. In MENU mode, select the TG CAL item. 
3. When  FULL  appears,  press  SELECT.    This  sets  the  current 
 
distance as the new Full setting. 
4. When EXIT appears, press SELECT return to Top Level Menu. 
 
13 