Regenerative Charge/Discharge Tester 69225/69212 Series User's Manual
Model
Channel
Charge Mode
Voltage Range
Maximum Current
Max Power
CC Mode Accuracy
Current Resolution
CV Mode Accuracy
Voltage Resolution
CP Mode Accuracy
Power Resolution
Discharge Mode
Voltage Range *
Maximum Current
Max Power
CC Mode Accuracy
Current Resolution
CV Mode Accuracy
Voltage Resolution
CP Mode Accuracy
Power Resolution
Measurement
Voltage Range
Voltage Accuracy
Voltage Resolution
Current Range
Current Accuracy
Current Resolution
Power Range
Power Accuracy
Power Resolution
Temperature Range
Temperature Accuracy
Temperature Resolution
1-4
69212-60-4
4
0-60Vdc
62.5A
1.25KW
0.1% stg.+ 0.05% F.S.
5mA
0.1% stg.+ 0.05% F.S.
2mV
0.2% stg.+ 0.1% F.S.
0.3W
0-60Vdc
62.5A
1.25KW
0.1% stg.+ 0.05% F.S.
5mA
0.1% stg.+ 0.05% F.S.
2mV
0.2% stg.+ 0.1% F.S.
0.3W
0-60Vdc
0.02% rdg.+ 0.02% F.S.
2mV
24A/62.5A
0.1% rdg.+ 0.05% rng.
5mA
1250W
0.12% rdg.+ 0.07% F.S.
0.3W
0-90°C
±2°C
0.1°C
69212-20-4
4
0-20Vdc
65A
1.25KW
0.1% stg.+ 0.05% F.S.
5mA
0.1% stg.+ 0.05% F.S.
0.5mV
0.2% stg.+ 0.1% F.S.
0.1W
0-20Vdc
65A
1.25KW
0.1% stg.+ 0.05% F.S.
5mA
0.1% stg.+ 0.05% F.S.
0.5mV
0.2% stg.+ 0.1% F.S.
0.1W
0-20Vdc
0.02% rdg.+ 0.02% F.S.
0.5mV
24A/65A
0.1% rdg.+ 0.05% rng.
5mA
1250W
0.12% rdg.+ 0.07% F.S.
0.1W
0-90°C
±2°C
0.1°C