Step 8:
Enter the correction value
If the correction value is greater than 5°C, then the correction value
is entered into the furnace by adding or subtracting accordingly to
the current value in the furnace. If the correction value is lower than
5°C, then no correction is needed.
Example 1:
Current calibration value of the furnace is 7
Value according to the conversion table is 6
New calibration value of the furnace is 13 (7+6)
Example 2:
Current calibration value of the furnace is -8
Value according to the conversion table is 7
New calibration value of the furnace is -1 (-8+7)
Step 9:
Complete the calibration
Calibration is now complete. Close the furnace head or select a program.
To return to the home screen, press either the [Home] button in the navigation bar or the Home key on the
membrane-sealed keypad.

2.2.4.4 Data backup

With the data backup function, individual programs and settings can be backed-up to a USB flash drive. We recommend
using this feature, e.g. before a software update or before sending in the furnace for maintenance purposes.
Moreover, furnace data saved on a USB flash drive can be restored on the furnace. However, restoring only works on the
same furnace with the identical serial number.
Data backup or data recovery requires only few steps:
Step 1:
Open data backup
Scroll to page 2 in the home screen and press the button
[Data backup].
2. Application
33