Delivery form

1 Handpiece holder,
1 Handpiece including power cord and power pack,
1 10 mm light probe,
3 Anti-glare cones,
1 Set of Operating Instructions

Accessories

The following accessories are available for bluephase C8:
REF.
Description
608537
Light probe 10 mm, black (G2)
608538
Pin-point light probe 6/2 mm,
black (G2)
551756
Anti-glare cones
592496
Anti-glare shield
627436
Handpiece holder C8 (G2)
607922
bluephase meter
608554
sleeves

Technical data

Operating voltage
5 VDC
Power Pack
100–240 VAC / 50-60 Hz /
max. 0.4 A
Output 5 VDC
TYP 15.2630
Manufacturer: Friwo

Product specifications

Operating conditions

Temperature
+10 °C to +40 °C
Relative humidity
30 % to 75 %
Ambient pressure
700 hPa to 1060 hPa
Dimensions of
handpiece holder
L=205mm; B=150mm; H=85mm
Weight of the
handpiece holder
190 g
Light source
polywave LED
Wavelength range
385–515 nm
Light intensity
800 mW/cm
Operating
6 min on / 4 min off
(intermittently)
Dimensions of
Handpiece:
L=260mm; B=42mm; H=120mm
Weight of the handpiece 200 g

Transportation and storage conditions:

Temperature
-20 °C to +60 °C (-4 to 140 °F)
Relative humidity
10 % to 75 %
Ambient pressure
500 hPa to 1060 hPa
bluephase C8 has to be stored in closed, roofed rooms.
Protect the device from severe jarring.