Product Specifications; Delivery Form; Technical Data - Ivoclar Vivadent Programat P710 Operating Instructions Manual

Hide thumbs Also See for Programat P710:
Table of Contents

Advertisement

9. Product Specifications

9.1 Delivery form

– Programat P710
– Power cord
– Vacuum hose
– Programat Firing Tray Set 2
– Automatic Temperature Checking Set ATK 2 (test set)
– USB download cable
– Programat WLAN Kit
– Programat Bluetooth Kit
– DSA Holder Kit
– var. accessories
Recommended accessories:
– Temperature Checking Set ATK 2
– Vacuum pump VP5, VP3 easy
9.2

Technical data

Power supply
Overvoltage category
Contamination level
Acceptable voltage fluctuations
Max. power consumption
Acceptable data for pumps of other manufacturers:
Max. output
Final vacuum
Electrical fuses
Dimensions of electrical fuses
Dimensions of the closed furnace
Usable size of the firing chamber
Max. firing temperature
Weight
86
110 – 120 V / 50 – 60 Hz
200 – 240 V / 50 – 60 Hz
II
2
± 10 %
14 A at 100 V
12 A at 110 – 120 V
8.5 A at 200 – 240 V
250 W /max. leakage current 0.75 mA
< 50 mbar
Only tested pumps should be used.
110 – 120 V:
125 V /T15 A (heating circuit)
125 V / T 5 A (vacuum pump)
200 – 240 V:
250 V / T8 A (heating circuit)
250 V / T 3.15 A (vacuum pump)
110 – 120 V:
Diameter 6.3 x 32 mm
200 – 240 V:
Diameter 5 x 20 mm
Depth: 495 mm
Width: 320 mm / 395 mm (with cooling tray)
Height: 320 mm
Diameter: 90 mm
Height: 80 mm
1200 °C (2192 °F)
18.7 kg

Hide quick links: