Firing With An Ivoclar Vivadent Program - Ivoclar Vivadent Programat CS6 Operating Instructions Manual

Hide thumbs Also See for Programat CS6:
Table of Contents

Advertisement

Loading for glaze and crystallization programs
For glaze and crystallization programs, place a maximum of four
crowns or one bridge on the IPS Speed Tray, depending on the
program. Only the IPS e.max CAD Crystallization Pin and
IPS Object Fix Putty/Flow may be used to secure the objects.
For more detailed information, please refer to the Instructions for
Use of the respective material.
Place the crystallization tray in the centre of the object tray in the
firing chamber using the special tweezers. If only one crown is
placed on the tray, the tray can be placed somewhat off the centre
so that the crown is located nearly in the centre to optimize the
crystallization and glazing result.
The crystallization tray must be placed in the centre of the object tray and within the circumferential groove in the
object tray.
The holes in the IPS Speed Tray are differently spaced so that a bridge can be ideally placed.
Predrying must not be deactivated for crystallization and glazing programs.
Remove the IPS Speed Tray from the furnace at the end of the glaze or crystallization program and place it on the
cooling tray for cooling.
2.3.2

Firing with an Ivoclar Vivadent program

Step 1:
Switch the furnace on
Switch on the furnace with the mains switch at the rear of the furnace.
The furnace conducts an automatic self-test.
Step 2:
Select the program
The program screen appears after the self-test has been completed.
Select the desired program using the [Arrow] buttons.
2. Application
49