Product Specifications; Delivery Form; Technical Data - Ivoclar Vivadent 747825BE Operating Instructions Manual

Table of Contents

Advertisement

9. Product Specifications

9.1 Delivery form

– Programat P310 G2
– Power cord
– Vacuum hose
– Programat Firing Tray Set 2
– Temperature Checking Set 2
– LAN download cable
– Operating Instructions
– Various accessories
Recommended accessories
– Temperature Checking Set 2
– Vacuum Pump VP5, VP3 easy

9.2 Technical data

Power supply
Overvoltage category
Contamination level
Acceptable voltage fluctuations
Max. power consumption
Acceptable data for pumps of other manufacturers:
Max. output:
Final vacuum:
Electrical fuses
Dimensions of electrical fuses
Dimensions of the closed furnace
Usable size of the firing chamber
Max. firing temperature
Weight
Safety notes
The furnace complies with the following standards:
IEC 61010-1:2010
IEC 61010-1:2010/AMD1:2016
EN 61010-1:2010/A1:2019
UL 61010-1:2012/R:2019-07
CAN/CSA-C22.2 No. 61010-1:2012/A1:2018-11
EN 61010-2-010:2020
IEC 61010-2-010:2019
UL 61010-2-010:2019
CAN/CSA-C22.2 No. 61010-2-010:2019
Radio protection, electromagnetic compatibility: EMC tested
110 – 120 V / 50 – 60 Hz
200 – 240 V / 50 – 60 Hz
II
2
± 10 %
12 A at 110 – 120 V
8 A at 200 – 240 V
250 W / max. leakage current 0.75 mA
< 50 mbar
Only tested pumps should be used
110 – 120 V:
125 V / T15A (heating circuit)
125 V / T5A (vacuum pump)
200 – 240 V:
250 V / T8A (heating circuit)
250 V / T3.15A (vacuum pump)
110 – 120 V: Diameter 6.3 x 32 mm
200 – 240 V: Diameter 5 x 20 mm
Depth: 465 mm
Width: 320 mm / 390 mm (with cooling tray)
Height: 320 mm
Diameter: 90 mm
Height: 80 mm
1200 °C
15.6 kg