Product Specifications; Delivery Form; Technical Data - Ivoclar Vivadent 747827AA Operating Instructions Manual

Table of Contents

Advertisement

9. Product Specifications

9.1 Delivery form

– Programat P510
– Power cord
– Vacuum hose
– Programat Firing Tray Set 2
– Automatic Temperature Checking Set 2 ATK2 (test set)
– USB download cable
– Operating Instructions
Recommended accessories
– Programat Accessories Set
– Temperature Checking Set ATK2
– Vacuum Pump VP5, VP3 easy
– Programat WLAN Kit

9.2 Technical data

Power supply
Overvoltage category
Contamination level
Acceptable voltage fluctuations
Max. power consumption
Acceptable data for pumps of other manufacturers:
Max. output:
Final vacuum:
Electrical fuses
Dimensions of electrical fuses
Dimensions of the closed furnace
Usable size of the firing chamber
Max. firing temperature
Weight
76
110 – 120 V / 50 – 60 Hz
200 – 240 V / 50 – 60 Hz
II
2
± 10 %
14 A at 100 V
12 A at 110 – 120 V
8.5 A at 200 – 240 V
250 W / max. leakage current 0.75 mA
< 50 mbar
Only tested pumps should be used.
110 – 120 V:
125 V / T15A (heating circuit)
125 V / T5A (vacuum pump)
200 – 240 V:
250 V / T8A (heating circuit)
250 V / T3.15A (vacuum pump)
110 – 120 V:
Diameter 6.3 x 32 mm
200 – 240 V:
Diameter 5 x 20 mm
Depth: 495 mm
Width: 320 mm / 395 mm (with cooling tray)
Height: 320 mm
Diameter: 90 mm
Height: 80 mm
1200 °C
18.7 kg

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Programat p510 g2Programat p510

Table of Contents