Ambient conditions for
operation
- Ambient temperature:
+5 °C to +40 °C
- Relative humidity of ambient air:
15% to 93% (non-condensing)
- Air pressure: 700 hPa to 1060 hPa
Users can operate the device both at
home and when out and about. Its use on
trains and aeroplanes is restricted to
passenger areas only. If using the device
in a car, it must be operated with battery
power.
Within professional healthcare facility
environments, the device must only be
used in in-patient wards and intensive
care units. The device must never be used
in areas exposed to elevated magnetic or
electrical radiation (e.g. in the vicinity of an
MRI scanner).
CAUTION:
When using portable wireless communi-
cation devices, you must ensure that
a distance of at least 30 cm is maintained
between all parts of the nebuliser system
and these devices (including their acces-
sories, e.g., antenna cables or external
antennas). Otherwise, the performance of
your nebuliser system may deteriorate.
CAUTION:
The controller must not be placed directly
beside or stacked on other devices for
operation. If operation near or stacked on
other devices is unavoidable, the control-
ler must be monitored continuously to
ensure proper operation.
®
eFlow
rapid - 2021-04
Ambient conditions for
transport and storage
- Temperature: -25 °C to +70 °C
- Relative humidity of ambient air:
0% to 93% (non-condensing)
- Air pressure: 500 hPa to 1060 hPa
These values for transport are valid for the
unpackaged product/ the unpackaged
aerosol head. The blister packing for the
aerosol head will only function properly
under the transport conditions described
on the system packaging.
NOTE:
Moisture from condensation can impair
the controller's functional capabilities.
Therefore, you should avoid extreme
temperature fluctuations. Wait until the
controller has reached room temperature
before beginning an inhalation session.
The batteries must be removed if you do
not expect to use the device for a lengthy
period.
Info:
Batteries vary considerably in terms
of storage life and output depending
on battery quality. The operating time
of about 90 minutes can only be
achieved with the batteries supplied
(4x or batteries with the same quality
and performance).