Maintenance
Maintenance
The following safety information has been added:
NOTE
Only perform maintenance measures if there is
no patient connected to the device.
The following table has been amended:
The following table shows the maintenance
intervals:
Component
CO
absorber
2
Water trap
Flow sensor
O
sensor
2
Internal lithium battery
Lead-gel battery
Cylinder pressure reducer for
high-pressure cylinders
Pressure reducer for PIN
1)
index
1) optional
40
Interval
If colored violet
As needed or if
soiled
As needed or if
calibration is no
longer possible.
As needed, if cali-
bration is no lon-
ger possible, or if
the O
SENSOR
2
FAIL ! message is
displayed.
Every 36 months
Every 3 years
After 6 years
1)
After 6 years
Measure
Replace
Replace
Clean/replace
Replace
Replace
Replace
Basic overhauling
Replace
Supplement Fabius plus SW 3.n
Responsible personnel
User
User
User
User
Experts
Experts
Experts
Experts