Alarm – Cause – Remedy
The terminology for the O2 sensor has changed.
Alarm
Alarm
priority
(8)
O
SENSOR FAIL !
2
(6)
O
SENSOR CAL DUE !
2
The causes and remedies for the following alarms have been
changed.
Alarm
Alarm
priority
(31)
INSP O
LOW !!!
2
(21)
FRESH GAS LOW !!
Supplement Fabius plus SW 3.n
Cause
The O
sensor was not cor-
2
rectly calibrated.
O
sensor capsule replaced
2
and/or the O
sensor not cali-
2
brated.
O
sensor capsule used up.
2
O
sensor not connected.
2
Sensor cable faulty.
More than 18 hours have
passed since the last O
sor calibration.
Cause
Inspiratory O
concentration
2
is below the lower alarm limit.
Incorrect setting of the switch
on the external fresh-gas out-
let.
Insufficient fresh-gas supply
in all ventilation modes.
Hose blocked/kinked.
Leakage in the breathing sys-
tem.
Hose not connected to the
breathing system.
Incorrect setting of the switch
on the external fresh-gas out-
let.
Troubleshooting
Remedy
Perform the procedure for
calibration of the O
(see chapter on calibrating
the O
sensor).
2
Perform the procedure for
calibration of the O
(see chapter on calibrating
the O
sensor).
2
Replace O
sensor capsule
2
and calibrate.
Connect the O
2
cable. Replace O
housing unit.
Perform the procedure for
sen-
calibration of the O
2
(see chapter on calibrating
the O
sensor).
2
Remedy
Check O
supply. Check
2
the setting of the flow con-
trol valve and lower O
alarm limit.
Check the setting of the
switch on the external
fresh-gas outlet.
Ensure sufficient fresh-gas
supply.
Check hoses.
Check breathing system.
Check the setting of the
switch on the external
fresh-gas outlet.
sensor
2
sensor
2
sensor
sensor
2
sensor
2
2
29