Dräger Fabius plus Supplement Manual page 31

Anesthesia workstation software 3.n
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Reprocessing
The chapter heading "Cleaning, Disinfection,and Sterilization"
has changed.
Disassembly
The following safety information has been added:
Information concerning dismounted
accessory parts and attached devices
Observe the instructions for use of the following
accessory parts:
Attached devices
– Endotracheal suction
– Hinged arms
– Monitors
– Sensors and cables
– IT systems
– AGS
– Workstation illumination
Removing the compact breathing system
The terminology for the O2 sensor has changed.
Before removing the compact breathing system,
the following hoses and cables must be removed:
– Flow sensor cable
– O
sensor cable and O
2
– Pressure measurement hose
– APL bypass hose
– PEEP/PMAX hose
Supplement Fabius plus SW 3.n
sensor
2
WARNING
Risk of infection
If no microbial filters are used, the breathing
system may become contaminated with
pathogens.
In this case, reprocess the breathing system
after every change of patient and insert a new
oxygen capsule.
The following safety information has changed:
CAUTION
Risk of damage to device and risk of injury
Reprocessing or sterilization can result in failure
of the O
sensor. Escaping substances can
2
damage the basic device and endanger the
patient and the user.
Do not reprocess or sterilize the O
Reprocessing
sensor.
2
31