Dräger Fabius plus Supplement Manual page 17

Anesthesia workstation software 3.n
Hide thumbs Also See for Fabius plus:
Table of Contents

Advertisement

Inserting a new O
The terminology for the O2 sensor has changed.
WARNING
Risk of electric shock
If the O
sensor capsule is replaced during
2
operation, leakage currents may be
transferred.
Do not touch the patient.
1 Remove the O
sensor (A) from the inspiratory
2
valve.
2 Unscrew the screw cap (C) from the O
housing.
3 Take the new O
sensor capsule from the
2
package.
4 Place the O
sensor capsule (B) in the O
2
sensor housing so that the ring-shaped
conductors touch the contacts in the O
housing.
5 Screw on the screw cap (C) tightly by hand.
6 Insert the O
sensor (A) back in the inspiratory
2
valve.
Supplement Fabius plus SW 3.n
sensor capsule
2
A
B
C
sensor
2
sensor
2
WARNING
Risk of incorrect O
An incorrectly fitted O
result in incorrect measurements.
– Make sure that the O
correctly inserted in the O
housing.
– Make sure that the O
inserted in the inspiratory valve.
2
Assembly and preparation
measurement
2
sensor capsule will
2
sensor capsule is
2
sensor
2
sensor is correctly
2
17