4.6.2 Cancel Calibration
Mode:
Key Path:
Dependencies and
Couplings:
Remote Command
Notes:
4.6.3 Store Calibration
Mode:
Key Path:
Dependencies and
Couplings:
Remote Command
Notes:
Chapter 4
Spectrum Analysis
Source, Open/Short CAL
Parameter is disabled if active mode is not Spectrum Analyzer
mode or a connection has not been established at you specified
IP address.
The Open/Short CAL can canceled at any time by pressing the
'Cancel' softkey on the softkey menu. If you cancel the
Open/Short Calibration at any point, previously stored
calibration data will not be overwritten
No SCPI. Front Panel only.
Spectrum Analysis
Source, Open/Short CAL, [Open CAL Form displayed], Continue,
[Short CAL Form displayed], Continue
This softkey is only displayed after an Open CAL and a Short
CAL have been completed.
The calibration procedure is completed by pressing this key.
When you press this key the stored open and short calibration
traces are averaged and copied to trace 3 (stored as the
reference trace), and you are returned to the Source menu.
The stored reference trace (trace 3) is cleared on a preset; if you
wish to re-use this calibration data you should save the trace
(File > Save > Type (Trace) > Save Now) and then recall it
following completion of the preset.
No SCPI. Front Panel only.
Source Key and Programming Commands
Open/Short CAL
65