therefore be slightly offset from the frequency limits of the source.
You can increase the number of sweep points or decrease the span of
the sweep to get closer to the frequency limit of your signal source,
but you will never be able to get right to the limit.
Frequency Limitations" on page 10
this limitation.
How do I view Signal Source errors?
Control does not check the error queue on your signal source. If you
have a problem with the signal source, for example, your sweep
failed to complete, check the error queue on your signal source for
information on any errors.
Signal Source IP Address Connection Problem
cross-over cable to connect the spectrum analyzer to the source
directly, then the IP address of both instruments must start with the
same first two integers. For example, if your analyzer IP address is
156.121.102.33, then your source IP address must be set to
156.121.xxx.xxx.