4.2 Amplitude
The amplitude parameter sets the power level and state of the external source, setting the state to ON
initializes the external source based on the current settings of the Spectrum Analyzer.
Mode:
Key Path:
Remote Command:
Factory Preset:
State Saved:
52
Spectrum Analysis
Source
:SOURce:EXTernal:POWer[:LEVel][:IMMediate][:AMPLitude]
<ampl>
:SOURce:EXTernal:POWer[:LEVel][:IMMediate][:AMPLitude]?
:OUTPut:EXTernal[:STATe] OFF|ON|1|0
:OUTPut:EXTernal[:STATe]?
−10.00 dBm, OFF
Saved in instrument state.