Agilent Technologies PSA Series Manual page 72

Spectrum analyzers
Hide thumbs Also See for PSA Series:
Table of Contents

Advertisement

Hints and Tips
FAQs and Helpful Hints and Tips
• Sweep Points = 101
• Offset = OFF
• Harmonic = OFF
This will cause the "Frequency band crossing not supported" error.
Reason:
The PSA band 0 stops at 3.05 GHz, and band 1 starts at 2.85 GHz. So
you must change your measurement to be completely in one of these
two bands. You can change:
• PSA start frequency > 2.85 GHz
or
• PSA stop frequency <3.05 GHz
Table 5-1
Frequency Bands in the PSA Series analyzers
(Microwave Preselector On)
Band
Number
0
1
2
3
4
5
6
Table 5-2
Frequency Bands in the PSA Series analyzers
(Microwave Preselector Off)
Band
Number
1
2
3
4
5
6
72
Frequency Range
3.0 Hz - 3.05 GHz
2.85 GHz - 6.6 GHz
6.2 GHz - 13.2 GHz
12.8 GHz - 19.2 GHz
18.7 GHz - 26.8 GHz
26.4 GHz - 31.15 GHz
31.0 GHz - 50.0 GHz
Frequency Range
3.0 GHz - 6.6 GHz
6.2 GHz - 13.2 GHz
12.8 GHz - 19.2 GHz
18.7 GHz - 26.5 GHz
26.4 GHz - 31.15 GHz
31.0 GHz - 41.0 GHz
Overlap
with Next
Band
200 MHz
400 MHz
400 MHz
500 MHz
400 MHz
150 MHz
N/A
Overlap
with Next
Band
400 MHz
400 MHz
500 MHz
100 MHz
150 MHz
500 MHz
Chapter 5