42
7-
Replacing the collimator and D814
6Collimator and D814
General
These instructions describe the work steps for replacing the collimator and D814.
Required documents
Software System Stand, SPB7-250.816.03....
System Start-up with WSH AWS, SPB7-250.815... for systems with AWS or
Installation Instructions, SPB7-250.812... for systems with DROC
QC Manuals
Tools
Standard tool kit
Service PC
Disassembling the collimator
WARNING
WARNING
MAMMOMAT Novation DR
Before you begin working on equipment, it is very important that
you disconnect it from the power supply at the main circuit break-
er.
Before removing or inserting any of the printed circuit
boards, switch off the equipment.
If the system is only switched off at the control panel, the line volt-
age will still be present at the generator line connection (see wiring
diagram).
A life-threatening hazard of electric shock exists.
SPB7-250.841.01.04.02
05.05
Collimator and D814
Page 42 of 64
CS SD 24
0
0
0
0
0
Siemens AG
Medical Solutions