Collimator N

Fig. 1:

Replacing the collimator

Unpack the collimator.
Remove the two colored screws from the transport bracket on the side of the blades
close to the focus.
Check the adjustment of the leaves close to the focus with the template.
Attach the collimator to the tube unit.
If plug X1 has a connector shell, unscrew it, cut it open, remove it.
Plug the X1 plug into the collimator and bolt it
Attach the cable at the panel using strain relief clamps
Perform the adjustment test when you replace the entire collimator.

Replacing electric components

Switch off the system
Replace the electric components according to the spare parts list and perform the ad-
justment test.
D1/S2.5
OFF
ON
LITHOSTAR Multiline
Strain relief
Collimator without prefiltration
Collimator with prefiltration
(On D1X7, check that the cable for the drive is connected)
RXL2-120.841.02.01.02
09.05
PC Board Replacement