Testing And Setting The Mas Counter - Siemens POLYMOBIL Plus Service Instructions Manual

Hide thumbs Also See for POLYMOBIL Plus:
Table of Contents

Advertisement

Measurements

Testing and setting the mAs counter

Unscrew the lid of the single tank cover.
Pull out the mAs jumper on D900 and connect the mAs meter.
Set 70 kV, 80 mAs on POLYMOBIL Plus and release exposure.
If the value displayed on the mAs meter does not coincide with the value set on
POLYMOBIL Plus, adjust the mAs counter with P1 on D915.
Release another exposure and compare the display.
WARNING
Siemens AG
Medical Engineering
Observe the cooling intervals for the tube.
SPR8-125.061.02
Rev. 01
Page 15 of 20
01.00
TD SD 24
5 - 15
5
POLYMOBIL Plus

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents