Checking The Mas Values - Siemens POLYMOBIL III Service Instructions Manual

Hide thumbs Also See for POLYMOBIL III:
Table of Contents

Advertisement

4 - 18
Fig. 2

Checking the mAs values

Remove the 4 Allen screws ( 3 mm key size) from the lid of the single tank and take off the
lid (Fig. 2).
Remove jumper "mA+ / mA - " in the single tank.
Connect the mAs meter to the "mA + / mA - " sockets (X038I-5/5).
Trigger the following exposures:
Setting at control panel
Remove the mAs meter and reinsert the jumper in the single tank at "mA+ / mA - " .
Reinstall the lid of the single tank.
POLYMOBIL III
40 kV, 200 mAs
66 kV, 100 mAs
90 kV, 1 mAs
Notes on trouble-shooting
valid mAs values
195 ... 205 mAs
97 ... 103 mAs
1,0 ... 1,3 mAs
RXB8-115.061.02
Rev. 01
10.95
Page 18 of 24
CS PS 24
4