4 - 18
Fig. 2
Checking the mAs values
•
Remove the 4 Allen screws ( 3 mm key size) from the lid of the single tank and take off the
lid (Fig. 2).
•
Remove jumper "mA+ / mA - " in the single tank.
•
Connect the mAs meter to the "mA + / mA - " sockets (X038I-5/5).
•
Trigger the following exposures:
Setting at control panel
•
Remove the mAs meter and reinsert the jumper in the single tank at "mA+ / mA - " .
•
Reinstall the lid of the single tank.
POLYMOBIL III
40 kV, 200 mAs
66 kV, 100 mAs
90 kV, 1 mAs
Notes on trouble-shooting
valid mAs values
195 ... 205 mAs
97 ... 103 mAs
1,0 ... 1,3 mAs
RXB8-115.061.02
Rev. 01
10.95
Page 18 of 24