Appendix C Using The Webserver - Stanford Research Systems SR865A Operation Manual

4 mhz dsp lock-in amplifier
Table of Contents

Advertisement

Appendix C
Using the Webserver
The SR865A webserver provides a quick start to interfacing via ethernet. Make sure to
set the IP address, subnet mask and gateway address for your network in the system
ethernet menu. Be sure to enable the webserver in the VXI-11 & Web settings. Press and
hold [Calc/system] to access the system menu.
To bring up the unit's webpage, simply type the unit's IP address into the browser bar of
a computer which can access the same subnet. Note that typical installations do not allow
the unit to be seen on the internet, only on the internal subnet. For full internet access,
consult with your IT administrator.
The homepage is shown below. Various instrument properties are displayed. This is an
easy way to check the ethernet settings without writing any software.
Click on the Interactive Control tab to display the command screen. Here you can try
sending commands or queries and checking the responses to SR865A commands. Click
on the SR865A Command List link to see the commands.