Recommended Monthly Maintenance

Warning
Wear rubber gloves, safety glasses, and protective clothing when doing any maintenance procedure. Report any
change in the operating condition of the processor to your service provider.
[1] Check that the "Daily Cleanup" section was completed as outlined on Page 25.
[2] Do up to Step 9 of the "Weekly Procedures" section outlined on Page 29.
[3] Do the "Replenishment Calibration Check Procedure" as outlined on Page 15.
[4] Drain all 3 Tanks by rotating the 3 Drain Valve Knobs
[5] Rinse the inside of all 3 Tanks with plenty of warm water.
[6] Clean any chemical residue from the Tanks by wiping the Tank walls with a clean, damp, lint-free cloth. To
prevent contamination, do not use the same cloth for the Fixer Tank and the Developer Tank.
[7] Close all 3 Drain Valves by rotating them
[8] On all Racks, check:
(a) that all Rollers rotate freely
(b) that the Rollers are clean and smooth
(c) that the Gears do not have excessive wear
(d) that the Bearings do not have excessive wear
(e) that the E-Rings and Springs are not broken, missing, or corroded
(f) that the Drive Chain on the Wash Rack does not have excessive wear and is adjusted correctly
(g) that the Dryer Rack is not dusty or dirty
(h) rinse the Air Tubes with water
[9] If necessary, install new parts.
[10] Check the Racks for squareness.
[11] Place Clamps on the Replenishment Tubes and clean the Replenishment Strainers located between the
Replenishment Pump and the Replenisher Tanks.
[12] Check the recommended Water Supply Filter. Replace it if necessary.
[13] Fill the Tanks. See Pages 8 through 14.
Caution
Install the Racks slowly to prevent contamination of the processing solutions.
[14] Install the parts in the following order:
1. Fixer Rack
2. Developer Rack
3. Wash Rack
4. Evaporation Covers
5. Dryer Rack
[15] Wipe any chemical deposits from the processing section of the processor, and clean the Film Sensors.
[16] Turn on the processor and check for correct operation by feeding a test film.
[17] Check the Drains for biological growth. If necessary, clean the Drains.