OPERATOR MANUAL
Weekly Preventive Maintenance Procedures
Warning
Wear rubber gloves, safety glasses, and protective clothing when doing any maintenance procedure. Report any
change in the operating condition of the processor to your service provider.
[1] Check that the "Daily Cleanup" section was completed as outlined.
[2] Remove the Top Cover, if it is not already removed.
Caution
• Handle these assemblies carefully to prevent changing their alignment.
• Do not clean the Racks or Rollers using abrasive materials.
• Do not wash the Rollers with water hotter than 37.5°C (100°F).
• Use the Splash Guard, Part No. 1C4019, between the Developer Tank and the Fixer Tank. Using the Splash
Guard will prevent contamination of the developer solution when you remove the Fixer Rack. Using the Drip
Tray, Part No.1C4011, when you remove or install any of the Racks will also help to prevent contamination of
the processing solutions.
Figure 17 Using the Splash Guard
Fixer Rack
26
Splash Guard
P/N 1C4019
Developer Tank
H130_0190HCB
H130_0190HA
September 1995 – 981089