Draining the Processor Tanks

Caution
If using a Silver Recovery Unit, disconnect the Drain Tube from the input side of the Silver Recovery Unit before
you open the Drain Valve. This will avoid flooding. Drain the fixer solution into a container or directly into the floor
drain. Check the local codes before draining solutions into a floor drain.
[1] Drain the Tank by rotating the Drain Valve Knob in either direction
place. When the Tank is drained, close the Drain Valve by rotating the Drain Valve Knob
direction.
Figure 5 Draining the Tanks
[2] Remove the Dryer Rack.
[3] Remove the Fixer and Developer Evaporation Covers.
[4] Remove the Fixer Rack and Developer Rack.
• To prevent solutions from one Tank spilling over into the next Tank, always:
– use the Splash Guard, Part No. 1C4019
– lower or raise the Racks very slowly when installing or removing them
• To prevent solutions from dripping into the processor or onto the Feed Tray:
– use the Drip Tray, Part No. 1C4011, when carrying Racks to and from the processor
– carefully tip the Racks when removing them from the processor to drain the remaining solution
– remove the Racks from the side of the processor, not from the front or back
[5] Rinse and thoroughly wipe the Racks and Evaporation Covers with a clean, lint-free cloth.
981089 – September 1995
Display
Panel
Replenishment Solutions
1
/
turn. You will feel the stem lock into
4
1
/
4
Feed Tray
Drain Valves
turn in either