Problem Solving

Diagnosing the Blink Patterns on the Display Panel

When the processor is operating, the condition of the processor is indicated by the 3 status indicators as outlined
below.
• When the Ready Indicator illuminates and neither the Wait Indicator nor Service Indicator is illuminated or
blinking, it indicates that:
– Both the developer solution and the Dryer are at the correct temperature.
– You may feed film.
• When the Ready Indicator illuminates and the Wait Indicator is blinking 3 times, it indicates that:
– The solutions are at the correct temperature.
– The Dryer has not reached setpoint temperature.
– Emergency films may be run, but may exit wet or damp.
• When the Wait Indicator illuminates and remains illuminated while the processor is operating, it indicates:
– A film is being fed and is not yet off the Feed Tray.
– The Film Sensors may be blocked by dirt or other debris.
• When the Wait Indicator blinks on and off while the processor is operating, it indicates that:
– There is a temporary error condition.
– There is an error condition that the operator may be able to correct. See Table 3 on Page 33.
• When the Service Indicator blinks on and off while the processor is operating, it indicates that:
– There is an error condition that the operator may be able to correct. The pattern of on and off flashes
indicates the error code. If the operator cannot correct the error condition, call for service.
– If film is fed into the processor during a service error, the processor will beep 2 seconds "On" then 2 seconds
"Off," while that film is processed.
• When the Service Indicator illuminates and remains illuminated while the processor is operating, it indicates
that:
– Service of the processor is required.
Table 2 Indicator Blink Patterns
Action
Feed Film
Feed Film
but may exit wet
Wait - Film in Process
Wait for the Film Feed Signal
Do Not Feed Film
981089 – September 1995
Ready
Wait
On
Off
On
Blinks
3 Times
On
On
Off
Blink or Off
Service
Off
Off
Off
Blink or Off
Problem Solving