Checking the exposure blocking

DLF switch on

Activate the DLF switch. In <Main menu> select <Configuration>, <Miscellaneous> and
<DLF switch>, press space to set switch to ON. Press <F2> to save.

Reduce grid speed

Reduce the grid speed. In <Main menu> select <Configuration> and <Grid speed>.
Set the following values:

Cassette loaded switch on

Set the cassette loaded switch on.
NOTE
NOTE
Siemens AG
Medical Solutions
Grid
Grid fast speed time
Grid fast speed
Grid slow speed
<ESC> to exit, <TAB> move to next entry field
1 Help
2 Save
3
The "cassette loaded" switch must be set to "ON" with the service
PC program.
The film cassette must be exchanged after each exposure. A sec-
ond exposure before exchanging the cassette is not possible
(exposure release is blocked).