Checks with high voltage

Checking X-ray tube high voltage, tube current and mAs values

1. System "ON".
2. Set the "cassette loaded" switch to "OFF" with the Service Program:
3. Trigger test exposures with the following exposure values and check the kV
4. Replace the jumper previously removed from sockets X3 and X4 on P.C. board
Siemens AG
Medical Solutions
-
Configuration
 Miscellaneous
values and tube current characteristic on the oscilloscope (see oscillograms).
The accuracy of the kV is
For mA values < 50 mA, the accuracy of the kV is
5 ms.
The tube current must rise quickly to the set value at the beginning of the expo-
sure and run linearly during the exposure.
The accuracy of the mAs product must be
D710.
Register 3
-
Cassette loaded check
5% plus
1.5 kV during the first 5 ms of the exposure.
&
&
10%.
&
SPB7-230.033.11
Rev. 06
05.06
.
5% plus
2 kV during the first
&
&
Page 5 of 10
M1000/3000 Nova
CS PS SP
Installation and Start-Up
10 - 5