Appendencies

Error codes

Symptom/mA signal
Constant current signal
or 1 = 0 mA
Unsteady current signal Undervoltage
I = 3.80 mA
I = 3.85 mA
I = 3.90 mA
I = 3.95 mA
I = 4.00 mA
I = 22.00 mA

Standard transmitter configuration

The transmitter is standard configured as described below on delivery:
pHix® Compact pH/temperatur. transmitter,  203160-4,6 - 203161-4,6 -
 203170-4,6 - 203171-4,6
Measuring range: 0-14 pH = 4-20 mA, 0-point = 4,6 pH
Temperature: 0-50° C/ 32-122° F = 4-20 mA
pHix® Compact pH/temperatur. transmitter,  203160-7,0 - 203161-7,0 -
 203170-7,0 - 203171-7,0
Measuring range: 0-14 pH = 4-20 mA, 0-point = 7,0 pH
Appendencies /Error codes
Fault
Electrical connection is
not made correctly
Undervoltage
Electrode error
Calibration not possible Exchange the electrode
pH lower than measuring
range
Error on startup
pH higher than
measuring range
pHix® Compact V3. NG (next generation)
Remedy
Check connection and
make the necessary
corrections
Check the voltage at
transmitter. Min. voltage @
max. load > 12 V DC
Check the voltage at
transmitter. Min. voltage @
max. load > 12 V DC
Exchange the electrode
Readjust the measuring
range to a lower range
Disconnect the transmitter
from PSU for min. 5
seconds and try again
Readjust the measuring
range to a higher range
EN 4.10 Version 2103