Xylem MJK pHix Compact V3 Manual page 60

Next generation
Table of Contents

Advertisement

Menu
Option
item
Device
pHix pH The sensor type
type
Electrode
pH
type
Zero
pH-
point 4.6
Electrode
pH
zero type
Zero
point 7.0
pH
Serial
number
Part
number
Sensor on-
chip serial
number
Number &
Revision
Compile
time
Slope
Zero point
pHix Redox
pHix Link software /Main configuration
Description
The pH electrode type
The sensor serial number
The sensor item number
Chip serial number
The current firmware in the sensor
The last time the firmware was changed
The slope is determined by measuring the difference
in mV readings pH buffers. The ideal slope should be
approximately -59.16 mV/pH at 25°C
The pH where the Zero Point is 0 mV, it means the
same amount of H+ and OH- ions.
pHix® Compact V3. NG (next generation)
EN 4.10 Version 2103
60