pHix® Compact V3. NG (next generation)
EN 4.10 Version 2103
pHix® Compact is mounted correctly when the electrode tip is emerged min.
½" in / 10 mm under the pipes inner pipe wall.

Eletrical mounting

pHix® Compact must not be installed in explosion hazardous
locations! pHix® Compact should be connected to an active 4-20 mA input or
supplied from a separate 10-28 VDC voltage source. Max power consumption
is 50 mW.
pHix® Compact 4-20 mA output(s) is galvanically separated from the liquids
potential.

Signal cable

Depending on the type, pHix® Compact comes with one or two galvanically
separated mA signal outputs and Modbus RS485. 
Note! pHix® Compact is delivered with cable made of PUR.
pHix® Compact with 2 x 4-20 mA outputs, one for pH or redox measurement
and one for temperature measurement.
Mounting /Eletrical mounting