Check that the connectors are absolutely clean and dry.
1.
Lubricate the O-ring in acid-free grease or Vaseline.
2.
Check that the O-ring is placed correctly in the recess of the new
electrode.
3.
Check that the hole for the electrode in the pHix® Compact is completely
dry – if necessary use a dry cloth.
4.
Screw in the new electrode and tighten by hand. Note! Do not use tools!
5. Finally, pHix® Compact must be
commissioning.

Mechanical mounting

pHix® Compact is made from Fortron (PPS) and/or stainless steel (EN 1.4404 /
AISL316L) and because of this, it has a tremendous resistance against
chemicals. pHix® Compact can, when mounted in a pipe, be used for
measuring in solvents and strong alcalic or acid liquids
Mounting /Mechanical mounting
pHix® Compact V3. NG (next generation)
buffer adjusted
 (see page 23)