pHix® Compact V3. NG (next generation)
EN 4.10 Version 2103
Buffer adjustment in pH 7 buffer. Buffer adjustment in pH 10 buffer.

Buffer adjustment using the tilt switch

Remember to clean the electrode as described earlier!
Only the built-in tilt switch is active when the switch ring has been removed for
more than 10 minutes.
The tilt switch inside pHix® Compact detects when the electrode tip is turned
upwards. When the electrode tip has been turned upwards for more than 30
seconds, pHix® Compact will go into buffer adjustment mode.
Proceed as described:
Remove pHix® Compact from the process, rinse the electrode, and turn the
electrode upside-down.
The buffer adjustment mode is initiated after app. 30 seconds - this is indicated
by the LED with 3 short flashes every 5 seconds.