Cleaning of the electrode

Use clean drinking water for cleaning the electrode.
A 10% hydrochloric acid solution can be used for cleaning if the electrodes are
greasy. Protein coatings are removed with Pepsin cleaning fluid.

Electrode lifetime

The expected lifetime of the electrodes is dependent of the measurement. The
lifetime is given with reservation as the physical conditions as well as the
temperature, pressure and pollution of the fluid has great influence on the
stated lifetime.
The lifetime is based on 77° F / 25° C, but if the temperature is doubled, the
lifetime is halved in worst case!
pH electrodes in purifying plants for ordinary household waste-water lasts
approximately one year, whereas the lifetime of pH electrodes in industrial
plants is dependent on measuring application and the process. Redox
electrodes have a lifetime of approximately two years, depending on cleaning
intervals.

Electrode spare parts

Item no. Description
160310
pHix® pH electrode, 0-point=4,6 pH
160311
pHix® pH electrode, 0-point=7,0 pH
160312
pHix® redox electrode
571030
Black rubber gasket for electrode
Union ring w. Viton gasket for mounting of pHix® on pipe - NOT
521409
AVAILABLE IN THE US
521441
Viton gasket for union ring - NOT AVAILABLE IN THE US
The following is recommended for cleaning and buffer adjustment:
Maintenance /Cleaning of the electrode
pHix® Compact V3. NG (next generation)
EN 4.10 Version 2103