Conducted dis-
3 V
0.15 MHz – 80
turbances
induced by RF
MHz, 6 V in
fields
ISM bands be-
IEC 61000-4-6
tween 0.15
MHz and 80
MHz, 80 % AM
at 1 kHz
3 V/m, 80 MHz
Radiated RF
– 2,7 GHz,
EM fields
80 % AM at 1
IEC 61000-4-3
kHz
See the RF
Proximity fields
wireless
from RF
communication
wireless
equipment
communication
table in
equipment
"Recommende
IEC 61000-4-3
d minimum
separation
distances"
Recommended minimum separation distances
Nowadays, many RF wireless equipments have being used in various healthcare locations
where medical equipment and/or systems are used. When they are used in close proximity to
medical equipment and/or systems, the medical equipment and/or systems' basic safety and
essential performance may be affected. The Fast-Fill has been tested with the immunity test
level in the below table and meet the related requirements of IEC 60601-1-2:2014. The customer
and/or user should help keep a minimum distance between RF wireless communications
equipments and the Fast-Fill as recommended below.
Test frequency
Band
(MHz)
(MHz)
385
380-390
Błąd! Użyj karty Narzędzia główne, aby zastosować 标题
Błąd! Użyj karty Narzędzia główne, aby zas
3 V
3V/m
Complies
Service
Modulation
Pulse
TETRA 400
modulation
18Hz
Portable and mobile RF
communications equipment should
be usedno closer to any part of the
Fast-Fill, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended minimum
separation distances
See the RF wireless communication
equipment table in "Recommended
minimum separation distances"
Maximum
Distance
power
(m)
(W)
1.8
0.3
Immunity
test
level
(V/m)
27
Page
34 / 37