8.Maintenance

8.1 Piston cleaning
Piston cleaning
Błąd! Użyj karty Narzędzia główne, aby zastosować 标题
1 5 0℃
Błąd! Użyj karty Narzędzia główne, aby zas
Heat to 150℃ to melt Gutta-percha, then
remove the cartridge.
NOTE
A small amount of Gutta-percha residue
have no effect of function.
Clean the residue inside the cartridge
needle with medical forceps.
Clean the overflow Gutta-percha on the top
of the handpiece with medical forceps.
When the medical forceps cannot get into
the cavity to clean, Heat the handpiece to
about 150℃, then unscrew the "Cleaning
Nut" with the bending tool.
NOTE
Heating to 150℃ to avoid damaging the
"Cleaning Nut".
Clean the inside of nut with medical
forceps.