Second Harmonic Distortion Test
The following test can be used to verify the combined input related spurious response of the FCN4760
Frequency Converter and Spectrum Master.
c. Equipment Required:
· Anritsu MG3692A Synthesized Signal Source with options 2A, 4 and 15A, or equivalent
· Anritsu 34RSN50 50 ohm adapter or equivalent
· Anritsu 15NNF50-1.5C RF Coaxial Cable or equivalent
· 10 MHz Reference Standard
· BNC male to BNC male coaxial cable
d. Procedure:
1. Connect the 10 MHz reference source to the Anritsu MG3692A Synthesized Signal Source.
2. Connect the output of the FCN4760 to the Spectrum Analyzer RF In port of the Spectrum Mas-
ter.
3. Connect the output of the source to the FCN4760 RF Input with the coaxial cable.
4. Connect the external power supply (Anritsu part number 40-168) to the Spectrum Master.
5. On the Spectrum Master, press and hold the
turn on the Spectrum Master. (This sets the instrument to the factory preset state.)
6. Turn on the 10 MHz reference source and the Anritsu MG3692A Synthesized Signal Source.
7. Set the MG3692A output to 5 GHz CW, with an RF output level of –45 dBm.
NOTE: Before continuing, allow a 30-minute warm up for the internal circuitry to stabilize.
8. Press the
AMPLITUDE
9. Enter –40 and press the
10. Press the
Scale
11. Press the
FREQ/SPAN
12. Enter
and press the
5
13. Press the
Span
14. Press the
BW/SWEEP
15. Press the
RBW Manual
the resolution bandwidth to 1 kHz.
16. Press the
VBW Manual
set the video bandwidth to 100 Hz.
17. Press the
MARKER
18. Select the
Edit
19. On the MG3692A Synthesized Signal Source, adjust the output level so that the M1 reading on
the Spectrum Master is –45 dBm at 5 GHz.
20. On the Spectrum Master, press the
21. Enter
5750 MHz
22. Press the
MARKER
23. Select the
Edit
24. Record the amplitude of the signal at M1 5750 MHz:
Second Harmonic Level @ 5750 MHz = ________ dBm
Spectrum Master MM
key and the
Ref Level
key to set the Reference Level to –40 dBm.
ENTER
soft key and enter 7 then press
key and then the
Center
soft key to set the center frequency to 5 GHz.
GHz
soft key and enter
. Press the
0.2
key and the
Bandwidth
soft key and use the Up/Down arrow key to select
soft key and use the Up/Down arrow key to select
key, then the
soft key.
M1
soft key. Enter
and press the
5
FREQ/SPAN
and press the
soft key to set the center frequency to 5750 MHz.
MHz
key and the
soft key.
M1
soft key. Enter
and press the
5750
key and press the
ESCAPE/CLEAR
soft key.
.
ENTER
soft key.
soft key to set the span to 0.200 MHz.
MHz
soft key.
soft key to set M1 to 5 GHz.
GHz
key and then the
Center
soft key to set M1 to 5750 MHz .
MHz
key to
ON/OFF
. Press
to set
1 kHz
ENTER
. Press
to
100 Hz
ENTER
soft key.
27