16. Record the calculated results in Table 3.
Table 3. Level Accuracy with Frequency Test Setting
Freq
Desired Sensor B reading for
(MHz)
0 dBm @ Attenuator output
30
550
1000
1244
1411
2000
2512
2850
2925
17. Using the power splitter, coaxial cable, adapters and fixed attenuator, connect the Spectrum Mas-
ter to the signal source and the power sensor as shown in Figure 2.
L i n e
O p e r a t e
S t a n d b y
Figure 2. Level Accuracy Test Setup
18. Connect the external power supply (Anritsu part number 40-168) to the Spectrum Master.
19. On the Spectrum Master, press and hold the
turn on the Spectrum Master. (This sets the instrument to the factory preset state.)
20. Press the
BW/SWEEP
Spectrum Master MM
ML2438A POWER METER
F r e q u e n c y
C l e a r
B a c k
E n t r y
S p a c e
L e v e l
O u t p u t
7
8
9
M o d u l a t i o n
O n O f f
4
5
6
S y s t e m
R F O u t p u t
5 0 9
1
2
3
.
0
+ / -
MG3692A SOURCE
34RSN50 ADAPTER
key and the
soft key.
Bandwidth
Desired Sensor B reading for
–39 dBm @ Attenuator output
MA2442A
15NNF50-1.5C
N241A50
10 dB
ATTENUATOR
34NN50A
ADAPTER
MT8212A
CellMaster
MODE
FREQ/DIST
AMPLITUDE
MEAS/DISP
S332D
key, then press the
ESCAPE/CLEAR
ESCAPE
1
2
CLEAR
START
AUTO
CAL
SCALE
3
4
SAVE
RECALL
SETUP
SETUP
5
6
LIMIT
MARKER
ENTER
7
8
SAVE
RECALL
RUN
DISPLAY
DISPLAY
HOLD
9
0
+
/
-
PRINT
ON
SYS
OFF
.
key to
ON/OFF
11