If Code has been accepted, the screen will display
If Code has not been accepted, the screen will display
If the security code has been changed and forgotten then contact the nearest K-TEK agent for override code.
To carry on with programming go to page 16.
6.0

INSTALLATION

The KSONIK MICRO is protected to IP65. The Transmitter is dust and waterproof so it can be mounted outside. The
KSONIK MICRO should be mounted using the bracket supplied. Do not install the KSONIK MICRO in areas of high vibration
as this may cause failure. Do not install the KSONIK MICRO in the close vicinity of electrical cable, SCR's or variable speed
drives. The installation of KSONIK MICRO is the most important section of this manual and has been divided up into 6 sub
sections.
1. The KSONIK MICRO must be fitted at least 0.50 m / 1.64 ft above the highest point of level.
2. Always use the plastic nut. The KSONIK MICRO must be fitted to a rigid support. Use mild steel or a suitable plastic.
Do not use stainless steel as this can cause ringing.
6 KSONIK MICRO Ultrasonic Level Transmitter | Operation instruction manual
Mode
Dist.
Level
Flow
Units: Meters
Empty Dist.:
10.000 m
Span:
10.000 m
Blanking:
0.500 m
Rate:
1.0 m/min
Application:
Liquids
Dist.
2.000 m
mA Output:
7.20 mA
Instant:
2.000 m
Temperature:
20°C
Percentage:
Steel bracket with
4 screws for mounting
2
20.00%
Wall