Onscreen Kscope - ABB KSONIK MICRO Operation & Instruction Manual

Ultrasonic level transmitter
Table of Contents

Advertisement

THE RELAY STATE SCREEN
ENTER
Press
while in run mode to get the RELAY STATE SCREEN.
2.
Number of cycles
1. Description of column
Relay
State
Hour
Cycles
5. Relay number
REL 1:
Off
LO
0
2
5. Relay state
3.
Hours relay has been on
1. Relay state headings.
2. The number of cycles the relay has been through.
3. The number of hours the relay has been on.
4. The state at which the relay is at ON, OFF, HI, LO and CO-Counter.
5. The relay indication number.

11.0 Onscreen KScope

ENTER
Press
twice while in run mode in order to get to the onscreen KScope.
1. The maximum returned echo size in volts e.g. 0.4V
2. Displayed instant distance from transducer to substance or object being measured. e.g.13.1 m
3. The Gain needed to get the particular returned echo signal to give a particular measurement. e.g. 50%
4. The amount of power needed to obtain an echo. e.g. 15% power
5. Maximum distance or span. e.g. 15.0 m
6. Graphical representation of the echo received by the transducer.
7. Threshold line whereby any echoes below this line will not be accepted.
8. Indication of good echo, noise, or loss of echo. e.g. E for good echo, N for noise or O for loss of echo.
9. The maximum voltage scale of the Onscreen KScope.
10. The echo loss timer started when no signal is present.
20 KSONIK MICRO Ultrasonic Level Transmitter | Operation instruction manual