Emc Notice - Barco Eonis MDRC-2222 WP User Manual

22-inch clinical display
Hide thumbs Also See for Eonis MDRC-2222 WP:
Table of Contents

Advertisement

Important information
Changes or modifications not expressly approved by the party responsible for compliance could void the
user's authority to operate the equipment.
FCC responsible: Barco Inc., 3059 Premiere Parkway Suite 400, 30097 Duluth GA, United States, Tel: +1
678 475 8000
Canadian notice
CAN ICES-3 (B)/NMB-3(B)
Brazilian local representative
BARCO LTDA, Av. Ibirapuera, 2332 – Torre II – 8° andar, Sao Paulo, Brazil

5.5 EMC notice

General information
This device is for use in professional healthcare facility environments only.
With the installation of the device, use only the delivered external cables and power supply or a spare part
provided by the legal manufacturer. Using another can result in a decrease of the immunity level of the device.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
WARNING: Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the MDRC-
2222, including cables specified by the manufacturer. Otherwise, degradation of the performance of
this equipment could result.
Electromagnetic emissions
The MDRC-2222 is intended for use in the electromagnetic environment specified below. The customer or the
user of the MDRC-2222 should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
This MDRC-2222 complies with appropriate medical EMC standards on emissions to, and interference from
surrounding equipment. Operation is subject to the following two conditions: (1) this device may not cause
1.
Active power for MDRC-2222 is less than 75 W
34
K5902146 (451920612625) /04
Compliance
Group 1
Class B
Not applicable
1
Complies
Eonis
Electromagnetic environment –
Guidance
The MDRC-2222 uses RF energy
only for its internal function.
Therefore, its RF emissions are
very low and are not likely to cause
any interference in nearby
electronic equipment.
The MDRC-2222 is suitable for use
in all establishments, including
domestic establishments and those
directly connected to the public low-
voltage power supply network that
supplies buildings used for
domestic purposes.