Emc Notice - Barco Nio 2MP LED User Manual

High-bright color display system
Table of Contents

Advertisement

Important information
Consult the dealer or an experienced radio/TV technician for help.
Changes or modifications not expressly approved by the party responsible for compliance could void the
user's authority to operate the equipment.
FCC responsible: Barco Inc., 3059 Premiere Parkway Suite 400, 30097 Duluth GA, United States, Tel: +1
678 475 8000
Canadian notice
CAN ICES-001(B) / NMB-001(B)
Brazilian local representative
BARCO LTDA, Av. Ibirapuera, 2332 – Torre II – 8° andar, Sao Paulo, Brazil

6.5 EMC notice

General information
This device is for use in professional healthcare facility environments only.
With the installation of the device, use only the delivered external cables and power supply or a spare part
provided by the legal manufacturer. Using another can result in a decrease of the immunity level of the device.
WARNING: Use of this equipment adjacent to or stacked with other equipment should be avoided
because it could result in improper operation. If such use is necessary, this equipment and the other
equipment should be observed to verify that they are operating normally.
WARNING: Use of accessories, transducers and cables other than those specified or provided by
the manufacturer of this equipment could result in increased electromagnetic emissions or
decreased electromagnetic immunity of this equipment and result in improper operation.
WARNING: Portable RF communications equipment (including peripherals such as antenna cables
and external antennas) should be used no closer than 30 cm (12 inches) to any part of the Nio 2MP
LED Display, including cables specified by the manufacturer. Otherwise, degradation of the
performance of this equipment could result.
Electromagnetic emissions
The Nio 2MP LED Display is intended for use in the electromagnetic environment specified below. The
customer or the user of the Nio 2MP LED Display should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/ flicker
emissions
IEC 61000-3-3
This Nio 2MP LED Display complies with appropriate medical EMC standards on emissions to, and
interference from surrounding equipment. Operation is subject to the following two conditions: (1) this device
may not cause harmful interference, and (2) this device must accept any interference received, including
interference that may cause undesired operation.
40
K5903058 /11
Nio 2MP LED Display
Compliance
Group 1
Class B
Class D
Complies
Electromagnetic environment –
Guidance
The Nio 2MP LED Display uses RF
energy only for its internal function.
Therefore, its RF emissions are
very low and are not likely to cause
any interference in nearby
electronic equipment.
The Nio 2MP LED Display is
suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.