Use of electrical surgical knives
Provide as much distance as possible between the electrosurgical generator and other electronic equipment
(such as monitors). An activated electrosurgical generator may cause interference with them and can disrupt
the functionality of the display.
Power connection – Equipment with internal power supply
This equipment must be earthed.
Power requirements: The equipment must be powered by the AC mains voltage.
The equipment is intended for continuous operation.
Transient over-voltage
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1
3rd. Ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
To maintain compliance with EMC Regulation, use only shielded interface cables for the connection to
peripheral devices.
Power cords
Europe: H05VV-F or H05VVH2-F PVC cord with appropriate EU plug.
US and Canada: "hospital grade" cord-set has to be used, provided with instructions to indicate that
grounding reliability can be achieved only when the equipment is connected to an equivalent receptacle
marked hospital only or hospital grade. These instructions need to be marked either on the equip. or on a
tag on the power cord
Do not overload wall outlets and extension cords as this may result in fire or electric shock.
Mains lead protection: Power cords should be routed so that they are not likely to be walked upon or
pinched by items placed upon or against them, paying particular attention to cords at plugs and
receptacles.
The power supply cord should be replaced by the designated operator only at all time.
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Korea: Use KC certified products; Plug: 250 V~, 16 A; Power cord: 60227 IEC 53, 3G0.75 mm² / 60227
IEC 53, 3G1.0 mm²; Connector: 250 V~, 10 A
Grounding reliability
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle.
Liquids and moisture
Never expose the monitor to liquids or moisture.
Never use the monitor near water - e.g. near a bathtub, washbasin, swimming pool, kitchen sink, laundry
tub or in a wet basement.
The equipment is IP20 compliant.
Moisture condensation
Do not use monitor under rapid temperature and humidity change condition or avoid cold air from air-
conditioning outlet directly.
Moisture may condense on the surface or inside of the unit, or create a mist residue inside the protection
plate, this is not a malfunction of the product itself, although it may cause damage to the monitor.
If condensation happens, let the monitor stand unplugged until there is no condensation.
Ventilation
Do not cover or block any ventilation openings in the cover of the set. When installing the device in a cupboard
or another enclosed location, heed the necessary space between the set and the sides of the cupboard.
Important information