Take the impact device in your hands. Place one hand
under the coil body (b). With the other hand make
a loading trigger movement; the impact body is then
released. Place the loading tube in the end position.
Carry out the following tasks regularly.
1. Clean the impact body of the impact devices P, PG
and U.
a ) Only the peg of the impact body
(marked with a circle, Fig. 69 ) may
be lightly oiled or greased.
b ) Remove dirt from the impact body
with a dry implement. Do not oil the
running surfaces
Fig. 69
2. Clean the guide tube (c) with the corresponding clean-
ing brush. Release or remove the support ring (a).
.
26

Accessories

.
Move the loading tube against the coil until the impact
body is freed. Clean the guide tube (c) and impact
body with a dry brush (do not oil).
3. After cleaning, push the impact body back into the
guide tube (c) until it is caught by the gripper.
4. Screw the support ring (a) onto the guide tube (c) up
to the end stop. The support ring is then correctly fitted.
The impact device is then once more ready for mea-
surement.
After complete assembly of the impact device, carry
out a functional check.
Connect the impact device to the display device.
Carry out 3 to 5 measurements on a paper roll.
The impact device is in correct working order if the
PAROTESTER2 does not display an error message.
12
Accessories
12.1
PAROLINK 3 data upload
Software
PAROLINK3 is a program for uploading measured data
from the PAROTESTER or PAROTESTER2 display
devices to an IBM or IBM-compatible PC via the RS232
interface.
PAROLINK3 can be used on the operating systems
WIN3X, WIN9.X, NT 4.0, WIN2000 and Windows XP.