Vega VEGAFLEX 82 Operating Instructions Manual page 30

Tdr sensor for continuous level measurement of bulk solids
Hide thumbs Also See for VEGAFLEX 82:
Table of Contents

Advertisement

6 Set up with the display and adjustment module
Medium, dielectric con-
stant
Max. adjustment level
Min. adjustment level
30
In this menu item, you can define the type of medium (product).
This menu item is only available if you have selected level measure-
ment under the menu item "Application".
You can choose between the following medium types:
Dielectric con-
Medium type
stant
> 3
Cereals, flour
1.5 ... 3
Granules, cement Lime, gypsum, cement
< 1.5
Dusts, wood chips Wood chips, sawdust
In this menu item, you can enter the max. adjustment for the level.
Adjust the requested percentage value with [+] and store with [OK].
Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. The distance refers to the sensor
reference plane (seal surface of the process fitting). Keep in mind that
the max. level must lie below the dead zone.
In this menu item, you can enter the min. adjustment for the level.
Adjust the requested percentage value with [+] and store with [OK].
Enter the suitable distance value in m for the empty vessel (e.g.
distance from the flange to the probe end) corresponding to the per-
Examples
All kind of cereals, wheat flour
VEGAFLEX 82 • Profibus PA

Advertisement

Table of Contents
loading

Table of Contents