VEGAFLEX 82 • Profibus PA
When using the probes without metal vessel wall, e.g. in plastic
vessels, the measured value can be influenced by strong electromag-
netic fields (emitted interference according to EN 61326: class A).
Use a probe in coax version for applications in liquids.
Fig. 3: Mounting in non-metallic vessel
1
Flange
2
Metal sheet
Concrete vessel
When mounting in thick concrete ceilings, VEGAFLEX 82 should be
mounted front flush to the lower edge. In concrete silos, the distance
to the wall should be at least 500 mm (20 in).
Fig. 4: Mounting in concrete silo
1
ø >160 mm
(ø >6.3")
4 Mounting
2
13