Units
Units (2)
Probe length
Application - Medium
type
Application
VEGAFLEX 82 • Profibus PA
In this menu item you select the distance unit and the temperature
unit.
For the distance units you can choose between m, mm and ft and for
the temperature units °C, °F and K.
In this menu item, you select the unit of the Secondary Value (SV2).
It can be selected from the distance units such as for example m, mm
and ft.
In this menu item you can enter the probe length or have the length
determined automatically by the sensor system.
When choosing "Yes", then the probe length will be determined
automatically. When choosing "No", you can enter the probe length
manually.
In this menu item you can select which type of medium you want to
measure. You can choose between liquid or bulk solid.
In this menu item you can select the application. You can choose
between metallic or non-metallic vessels.
Note:
The selection of the application has a considerable influence on all
other menu items. Keep in mind that as you continue with the param-
eter adjustment, individual menu items are only optionally available.
You have the option of choosing the demonstration mode. This mode
is only suitable for test and demonstration purposes. In this mode, the
sensor ignores the parameters of the application and reacts immedi-
ately to any change.
6 Set up with the display and adjustment module
29