Directing The Camera - Dentsply Sirona CEREC AF Operating Instructions Manual

Hide thumbs Also See for CEREC AF:
Table of Contents

Advertisement

Available languages

Available languages

7 Operation

7.4 Directing the camera

48
Directing the camera
7.4
CAUTION
After each use
Reprocess the camera after each patient.
➢ Follow the instructions in chapter „CEREC Omnicam  [→ 57]" in
order to avoid cross-contamination between patients.
The CEREC Omnicam acquires images which are used during the
ongoing measurement in spatial relation to each other (image
registration).
During the acquisition and then during the ongoing registration process,
a distinctive sound can be heard.
If the registration cannot be implemented, the acquisition flow is
suspended. You are informed of this by means of a sound. This is
different to the sound emitted during successful acquisition. You can
adjust the volume in configuration.
IMPORTANT
Registration error
Should a registration error occur, you must return to another acquired
point.
To start with, practice this procedure on the model and then on
intraoral areas.
➢ Move the CEREC Omnicam to a position where a successful
acquisition was taken. A point that has already been acquired in
the occlusal area is best.
Ä You will be able to hear the sound for registered acquisitions.
➢ Continue the acquisition.
Divide the acquisition into four consecutive sequences:
1. Occlusal
2. Buccal
3. Lingual
4. Proximal
Dentsply Sirona
Operating Instructions (valid for Canada)
66 37 487 D3652
D3652.201.01.21.02    03.2021

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cerec af connectCerec aiCerec ai connectCerec ai for rear

Table of Contents