Dentsply Sirona
Operating Instructions LEDview Plus
65 67 510 D3642
D3642.201.01.03.02 01.2018
Dismounting the patient mirror
1. Turn the patient mirror 45° counterclockwise.
2. Remove the patient mirror from the front.
4.5.3

Camera holder

General information on the camera holder
The accessory interface must be installed on the operating light in order
to use the camera holder.
You can install the camera holder with the camera mounted on the
mounting adapter of the accessory interface and remove it again when
not in use.
Standard commercial cameras with tripod thread can be fastened to the
camera holder. Please note that there is no camera included in the
camera holder retrofit kit and this is not supplied by Dentsply Sirona.
The total weight of the camera may not exceed 240 g.
The camera and the power unit that supplies the camera must at a
minimum fulfill one of the following standards:
● IEC 60601-1 (2MOOP)
● IEC 60950-1
● IEC 62368-1
The camera must be capable of being disinfected in order to meet the
hygiene requirements.
The dimensions for the camera must be noted when selecting the
camera. The mounted camera must not restrict the operating light's full
range of motion.
IMPORTANT
The camera can conceal the keyboard at the camera holder with the
ceiling model.
Select the size for the camera on the ceiling model in such a way that
both of the outer keys of the keyboard (On/Off and Composite) can still
be used. Remove the camera holder from the operating light in order to
operate both of the inner keys if these are not accessible.
4 Operation
4.5 Accessories for the operating light
37